From the Translational Cardiac Electrophysiology Laboratory (Y.O., B.D.H., S.B.J., T.J.B., C.J.O., D.L.P.), Division of Cardiovascular Disease, Department of Internal Medicine, Health Sciences Research (D.O.H.), Mayo Clinic, Rochester, Minn.; and Biosense Webster (A.A., A.G.), Diamond Bar, Calif.
Correspondence to Douglas L. Packer, MD, 2-416 Alfred Building, Saint Marys Hospital Complex, Mayo Foundation, Rochester, MN 55902. E-mail packer.douglas{at}mayo.edu
Received September 18, 2007; accepted April 7, 2008.
| Abstract |
|---|
|
|
|---|
Methods and Results— Twelve dogs underwent evaluation using a newly developed Carto-based 3D ultrasound system. After fiducial clip markers were percutaneously implanted at critical locations in each cardiac chamber, 3D ultrasound geometries, derived from a family of 2D intracardiac echocardiographic images, were constructed. Point-source error of 3D ultrasound-derived geometries, assessed by actual real-time 2D intracardiac echocardiographic clip sites, was 2.1±1.1 mm for atrial and 2.4±1.2 mm for ventricular sites. These errors were significantly less than the variance on CartoMerge computed tomographic images (atria: 3.3±1.6 mm; ventricles: 4.8±2.0 mm; P<0.001 for both). Target ablation at each clip, guided only by 3D ultrasound-derived geometry, resulted in lesions within 1.1±1.1 mm of the actual clips. Pulmonary vein ablation guided by 3D ultrasound-derived geometry resulted in circumferential ablative lesions. Mapping in 15 patients produced modestly smaller 3D ultrasound versus electroanatomic map left atrial volumes (98±24 cm3 versus 109±25 cm3, P<0.05). Three-dimensional ultrasound-guided pulmonary vein isolation and linear ablation in these patients were successfully performed with confirmation of pulmonary vein entrance/exit block.
Conclusions— These data demonstrate that 3D ultrasound images seamlessly yield anatomically accurate chamber geometries. Image volumes from the ultrasound system are more accurate than possible with CartoMerge computed tomographic imaging. This clinical study also demonstrates the initial feasibility of this guidance system for ablation in patients with atrial fibrillation.
Key Words: ablation mapping 3D ultrasound imaging
| Introduction |
|---|
|
|
|---|
Although displaying an ablation catheter and integrating its position within CT/magnetic resonance geometry is possible, the accuracy of the integrated 3D geometry is highly dependent on image quality and, more importantly, on the merge process.1–8 Errors may stem from changes in cardiac chamber volume or rhythm occurring between the time of 3D anatomic image acquisition and electroanatomic mapping in the electrophysiology laboratory. Gating of image acquisition to the cardiac cycle and respiratory phase may also affect rendered volumes, as may anesthesia administered in the electrophysiology laboratory.
Clinical Perspective see p 120
Alternatively, techniques that provide high-definition real-time anatomic images should mitigate those factors contributing to matching errors. Recently, an ultrasound-based 3D imaging modality has been developed to guide catheter navigation. This system creates anatomically accurate volumes from real-time 2D intracardiac echocardiographic (ICE) images facilitating interventional navigation within target cardiac chambers. Nevertheless, the accuracy and utility of this system have not been established. This study was therefore undertaken to (1) validate the accuracy of integrated 3D ultrasound imaging, (2) compare the accuracy of catheter manipulation using this approach with that of merged CT image–based navigation, and (3) assess the clinical utility of 3D ultrasound image–guided intervention in patients with atrial fibrillation (AF).
| Methods |
|---|
|
|
|---|
Animal Validation Study
Twelve mongrel dogs (30 to 40 kg) were studied according to a protocol approved by the Mayo Foundation Institutional Animal Care and Use Committee, with general methods described previously.9 After deep anesthesia was established, positive-pressure ventilation (rate 14/min, inspiration:expiration=1:2) instituted, vascular access was percutaneously obtained. Surface ECGs and blood pressure were continuously monitored. Transseptal catheterization was accomplished under fluoroscopic and phased array ICE (10F, 64-element, 5.5 to 10.0 MHz, Siemens) guidance.
Clip Marker Implantation
To support the subsequent validation studies, 6 to 9 intracardiac anatomic marker clips (QuickClip 2, 8 mmx2 mm) were percutaneously implanted under fluoroscopic and 2D ICE guidance at specific anatomic locations including the pulmonary vein (PV) orifices, left atrial appendage, superior vena cava–right atrial (RA) junction, RA appendage, and apex or lateral wall of the right ventricle (RV) and left ventricle (LV). These were well visualized by ICE in each animal, allowing their subsequent use as consistent fiducial points for segmentation, registration, and image fusion studies as required throughout the entire validation study.
CT Imaging and Image Segmentation
Thereafter, a 64-slice multidetector CT scan (Siemens) was acquired. Helical CT scanning (330 ms/rotation, 0.75 slices/0.4-mm interval, 0.2 pitch) was performed 30 seconds after intravenous injection of 100 mL of contrast media at a rate of 4 mL/s. Volume image acquisitions were gated at 80% of the R-R interval for atrial and at 0% of the R-R interval (R wave) for ventricular imaging during the end-expiratory phase.
The axial CT images ware transferred to the electroanatomic mapping system equipped with CartoMerge Image Integration Software (Biosense Webster). The surface reconstruction of the RA with superior vena cava, left atrium, PVs, RV, and LV were segmented from each chamber volume through a 3-step process as described previously.1–5,7,8 Individual implanted clips within each chamber were also segmented to serve as specific, predetermined 3D CT anatomic fiducial points.
3D Volume Rendering With Intracardiac Ultrasound
An electroanatomic location sensor was incorporated into the ICE catheter tip to allow the establishment of (1) tip location in 3D space, (2) tip direction, and (3) the origin and direction of the 2D ICE sector image. Ultrasound images were relayed to the CARTO XP system (CartoSound) and displayed within the same coordinate system as the subsequently generated electroanatomic maps. Three-second segments of 2D ultrasound images were acquired during ECG gating to a coronary sinus (CS) atrial electrogram for atrial imaging and to the R wave of V1–3 for ventricular imaging. Although the 2D ICE images were not automatically gated to respiration by the system, images used in the analysis were acquired in late expiration (79%) or did not show appreciable variability between inspiration and expiration (21%). A single gated image was displayed on the systems resident "Ultrasound Viewer," from which the endocardial surface contour was manually drawn on the basis of echointensities at the blood and tissue interface (Figure 1A). Each surface contour was thereby stored as a series of individual electroanatomic mapping points delineating each planar anatomic shape. By repeating this process, a family of chamber contours was acquired, collated into a complete volume rendering using point–to-point component interpolation (Figure 1B and 1C), and displayed on the Map Viewer (Figure 1D). Each target 3D ultrasound chamber was acquired during a rotational sweep of the chamber. In some cases, several catheter rotations, further ICE catheter insertion or withdrawal, or imaging head angulation was required to complete the target 3D geometry. In addition, the ICE imaging venue was specific for a given chamber; ie, an RA venue was used for LA imaging and RA and RV venues were used for RV and LV imaging. Subsequently, in a separate process, the "tag points" reflecting each implanted clip "insertion site" were manually drawn from repeated ICE images onto the 3D ultrasound–derived geometries to serve as "blindable" anatomic surface markers for use in validation studies (pink-unfilled circle) (Figure 2A). Other electroanatomic points were also acquired by using a separate mapping catheter and registered to the reconstructed 3D ultrasound volume, all registered within the same coordinate system as the ICE images (Figure 1D).
|
|
For the second method ("outside in" validation, Figure 3), the opposite approach was utilized. The mapping catheter was placed at an apparent clip "tag point" (pink-unfilled circle) on the 3D ultrasound-derived geometry, while the operator was blinded to the real-time 2D ICE/fluoroscopic images, and this apparent clip "tag point" of origin was marked with a yellow-filled circle (Figure 3A). This tag point was then compared with the corresponding actual clip insertion site established by repositioning the Carto mapping catheter under direct 2D ICE/fluoroscopic guidance (blue-filled circle). The position error of the "outside in" validation was defined as the shortest distance from this 3D ultrasound-based tag point origin to the corresponding 2D ICE/fluoro-based clip insertion site (yellow-filled circle versus blue-filled circle) (Figure 3B).
|
|
The utility of 3D ultrasound–derived geometry for guiding linear ablation (each ablation: 30 seconds, 50 W, 60 °C) was also examined. Creation of a mitral isthmus line was guided in 6 dogs by both LA and LV volumes. Thereafter, circumferential PV ablation (right superior PV [n=8] and superior branch of left superior PV [n=8]) was guided by 3D ultrasound–derived LA and PV geometries. Subsequently, after VF induction with high-rate burst pacing and animal exsanguination all under deep anesthesia, the relationship between implanted clip markers and ablative lesions was evaluated. The position error was based on 2 measurements (Figure 5): (1) distance between the center of the clip insertion point and the center of the lesion and (2) the distance between the center of the clip insertion point and the edge of the lesion.
|
Clinical Study of 3D Ultrasound Image Accuracy and Guided Ablation
Fifteen AF patients also underwent clinical electrophysiological evaluation according to previously described methods.9 These patients were identified from an institutional review board–approved database study. An Authorization to Use and Disclose Protected Health Information form was on file before retrospective analysis of this prospective cohort. After vascular access was obtained, catheters for recording and pacing were positioned in the RA, RV, and CS. A 10F ICE catheter was positioned via the right femoral vein and advanced into the RA. Double transseptal catheterization was performed under ICE guidance and mapping and Lasso catheters advanced into the LA. Intravenous heparin was given to maintain an activated clotting time >300 seconds during the procedure.
The 3D ultrasound images of the LA and PVs were acquired and processed according to the same methods as deployed in the canine studies. These were compared with acquired standard electroanatomic mapping datasets as previously described.1–5,7,8 The outside-in validation was also performed as described in the animal validation methods. In brief, 4 anatomic points (venoatrial junction of each PV) on the 3D ultrasound geometry and subsequent electroanatomic mapping were marked with yellow-filled circles. These tag points were then compared with the fiducial corresponding anatomic sites (blue-filled circle) based on an independent 2D ICE slice.
In all patients, wide area circumferential ablation around each pair of ipsilateral PVs was undertaken with the use of an open irrigated 7F, 3.5-mm-tip catheter. Radiofrequency energy was delivered (each ablation: 15 to 30 s, 25 to 40 W, 40°C, irrigation flow 17 to 30 mL/min) until PV isolation was accomplished. Catheter navigation was guided by 3D ultrasound imaging with limited use of 2D ICE/fluoroscopic images. If venoatrial electric disconnection was not achieved with the wide area circumferential ablation, additional Lasso-guided segmental ostial ablation, targeting sites of electric breakthroughs, was performed.
Statistical Analysis
All statistical analyses were performed with SAS 9.1.3. Continuous variables were presented as mean±SD, and comparisons between groups were completed by using 2-sample t test. Because remaining analysis included multiple observations from the same dog or patient, the Generalized Estimating Equation models were used to adjust for potential effects of repeated measurements in each dog. Values of P<0.05 were considered statistically significant. In the setting of multiple comparisons, the Bonferroni method was also used, with a P<0.008 taken as significant.
| Results |
|---|
|
|
|---|
Baseline Validation Study
With 96 registered clips used for analysis, the overall position error between the clip insertion sites, as established by 2D ICE/fluoro–guided electroanatomic navigation (gold standard) and clip site tag points annotated on the 3D ultrasound–derived geometry (inside-out validation), was 2.0±1.5 (range 0 to 6.2) mm. For each chamber, the error was 2.3±1.2 (range 0 to 4.4) mm in the RA, 1.9±1.4 (range 0 to 4.5) mm in the LA, 1.9±1.4 (range 0 to 4.8) mm in the RV, and 2.0±1.9 (range 0 to 6.2) mm in the LV (P=NS between cardiac chambers, Table 1).
|
Ablation Accuracy Assessment
All ablative lesions targeting the implanted 23 RA and 24 LA clips were seen on gross pathology. Twenty-six of the 27 clips in the RV and 18 of the 22 clips in the LV were also used for guidance validation. Radiofrequency energy delivery–created lesions, which were 4.6±1.6 mm wide and 6.1±2.1 mm long, showed overall position errors of 1.1±1.1 mm (range 0 to 4.3 mm) from the center and 0.5±0.9 mm (range 0 to 4.0 mm) from the edge of the ablative lesions. For the individual cardiac chambers, position errors for the center versus edge of the lesions were 1.1±1.1 (range 0 to 3.5)/0.5±0.8 (range 0 to 3.0) mm RA lesions, 1.3±1.3 (range 0 to 4.3)/0.8±1.1 (range 0 to 4.0) mm LA ablation sites, and were 0.9±1.1 (range 0 to 3.0)/0.4±0.7 (range 0 to 2.2) mm in RV and 1.1±0.9 (range 0 to 3.5)/0.1±0.3 (range 0 to 1.0) mm in LV lesions (P=NS between cardiac chambers for the lesion center, Figure 5, Table 2).
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Validation Study of 3D Ultrasound–Derived Geometry
Two-dimensional ICE imaging is commonly used to precisely display real-time anatomic information.7–9,11,12 In the clinical electrophysiological environment, such imaging is used to monitor (1) catheter positioning, (2) catheter tip/tissue orientation, (3) catheter tip/tissue contact, (4) lesion formation, and (5) microbubble development. The real-time characteristics of the 2D ICE image for detection of cardiac surface contours have prompted the development of this novel integrated 3D ultrasound system. Because differences between real-time 2D ICE-based surfaces and the resulting 3D ultrasound rendering may limit image accuracy, validation of this approach was required before its widespread clinical application.
In both animal and human studies, the 3D ultrasound volume renderings showed an acceptable point-source position error of 2 mm, a component of which included electromagnetic location error of <0.8 mm inherent in the system.13 Factors related to respiratory cycle may also contribute to observed error. Recent studies have shown significant differences in cardiac chamber dimensions occurring over the course of expiration and inspiration.14,15 Although such differences were seen during this study, their impact was minimized by comparing 3D ultrasound and CT volumes acquired during the expiratory phase, even though the system did not automatically gate for respiratory cycling. Furthermore, this study showed
1 mm resulting error in ablative lesion location, supporting the feasibility and acceptability of this approach to 3D navigation. The size of ablative lesions in this study and their precise positioning suggest this error is not limiting in the radiofrequency ablation arena.
Comparison of the Accuracy of Merged CT and 3D Ultrasound–Derived Images
These data confirm the discrepancies in merged CT/magnetic resonance images reported in recent studies.1–5,7,8 Several prior integration studies2–4 showed a match error of <3 mm as calculated from average global surface match of the registered CT geometry and 3D electroanatomic mapping. Nevertheless, all clinically acquired electroanatomic mapping points used in such a process of surface merging are attached to the closest 3D geometry surface without regard to any prespecified landmark pairs.2–5,7,8 Therefore, average global match error would artificially appear small, masking potential single-point error at specific target sites.
In contrast, more recent validation studies7,8 focusing on position error at individual target sites showed >5-mm inaccuracies despite favorable average global surface match. The present study confirms that the surface merge alignment with CT-generated images results in larger point-source errors (5.4±2.5 mm for atria and 6.6±3.1 mm for ventricles) as established from the preimplanted fiducial clips, although chamber geometries seemed better aligned, as suggested by the smaller average surface mismatch. Nonetheless, despite manual adjustments, the point-source position error remained larger with CartoMerge guidance than seen using 3D ultrasound geometries.
Such persistent position errors after adequate adjustment of the merge process are likely due to the presence of different chamber conformations in CT and 3D electroanatomic data sets. This can occur because of various LA chamber status changes occurring in the time interval between CT image acquisition and intervention, including changes in volume, rhythm, cardiac cycle, or respiration. Also, the endocardial surface detection of CT/magnetic resonance imaging is achieved by using contrast material and completed by a limited semiautomatic segmentation process. Poor CT image quality at faster heart rates may further increase inaccuracy, particularly during ventricular imaging.16 In contrast, the 3D ultrasound image volumes have the advantage of real-time confirmation by combined 2D ICE imaging and electroanatomic measurements.
Anatomy-Based Ablation Guided by 3D Ultrasound–Derived Geometry
This study shows the accuracy of 3D ultrasound–guided ablation, which was successful in creating continuous lesions even in the setting of complex anatomy within the small canine LA. As expected, this 3D ultrasound geometry–based approach was also feasible with a variety of AF ablation methods and patient characteristics. The accuracy and feasibility of this approach were confirmed through the validation studies and the qualitative configuration of ablative lesion sets in dogs and the comparisons of the 3D ultrasound geometries and electroanatomic mapping and establishment of PV-antrum entrance block in humans. The clinical data also demonstrated that electroanatomic maps of the LA tended to be larger than 3D ultrasound volumes. This may be partially explained by tissue surface distortions created by manual catheter manipulation. Importantly, the PV orifice was more easily and accurately detectable with 3D ultrasound geometry, as reflected by larger PV orifice areas and smaller errors than seen with electroanatomic mapping. This should allow for more accurate navigation around the PVs, thereby improving the efficacy and safety of AF ablation. Consistent with this, we did not observe any complications such as tamponade, cardiac perforation, or injury to the esophagus or phrenic nerve in the animal and clinical studies. More completely establishing safety and efficacy will require additional larger comparative studies, however.
Study Limitations
There are several technical limitations of this investigation. First, the current system did not allow automatic gating to respiratory cycle. Nevertheless, 79% of images used in the analysis were acquired in late-expiratory phase, and the remaining clips showed no appreciable perimeter changes between inspiration and expiration, thus minimizing the potential impact of this factor. In fact, respiratory cycling made a very small contribution to observed error. Second, the resolution of 2D ICE imaging is less than that of the CT imaging. Possible factors such as acoustic shadowing or incomplete penetration could limit the ability to complete 3D ultrasound rendering. Nevertheless, a prior study from this laboratory showed excellent resolution of tissue structures by 2D ICE imaging at a distance of 0.5 to 8 cm from the tip.10 Multiple catheter rotations, further ICE catheter insertion or withdrawal, or imaging head angulation also enabled the completion of the 3D ultrasound geometry in all animal and human cases. Third, the 3D ultrasound rendering does not establish catheter tip/tissue contact, except as reflected by the catheter tip color. Because the 3D ultrasound geometry consists of a family of the outlined contours derived from the 2D images, catheter tips at or beyond the 3D ultrasound geometry will still disclose tissue contact, as will the accompanying ICE images. Despite these issues, the 3D ultrasound system provided a robust basis for mapping and ablation, with little inherent error from technical factors, with further feasibility of application demonstrated in the clinical studies.
| Acknowledgments |
|---|
This study was supported in part by an unrestricted research grant from Biosense Webster, Inc.
Disclosures
Dr Packer received other clinical and animal research support from Biosense Webster. during the conduct of this study. Dr Packer has also received research grants from Hansen Medical, St. Jude Medical, Boston Scientific, ProRhythm, CryoCath Technologies, and CardioFocus; has received honoraria from Bard EP, CryoCath Technologies, Medtronic, Biosense Webster, Hansen Medical, and St. Jude Medical; has royalty interest in St. Jude Medical; and has served as a consultant to or on the advisory boards of Biosense Webster, Bard EP, Boston Scientific, CryoCath Technologies, Hansen Medical, and SanofiAventis. A. Altman and A. Govari are employees of Biosense Webster. Ms. Johnson receives royalty payment from St Jude Medical. The other authors report no conflicts.
| References |
|---|
|
|
|---|
2. Tops LF, Bax JJ, Zeppenfeld K, Jongbloed MR, Lamb HJ, van der Wall EE, Schalij MJ. Fusion of multislice computed tomography imaging with three-dimensional electroanatomic mapping to guide radiofrequency catheter ablation procedures. Heart Rhythm. 2005; 2: 1076–1081.[CrossRef][Medline]
3. Dong J, Dickfeld T, Dalal D, Cheema A, Vasamreddy CR, Henrikson CA, Marine JE, Halperin HR, Berger RD, Lima JA, Bluemke DA, Calkins H. Initial experience in the use of integrated electroanatomic mapping with three-dimensional MR/CT images to guide catheter ablation of atrial fibrillation. J Cardiovasc Electrophysiol. 2006; 17: 459–466.[CrossRef][Medline]
4. Kistler PM, Rajappan K, Jahngir M, Earley MJ, Harris S, Abrams D, Gupta D, Liew R, Ellis S, Sporton SC, Schilling RJ. The impact of CT image integration into an electroanatomic mapping system on clinical outcomes of catheter ablation of atrial fibrillation. J Cardiovasc Electrophysiol. 2006; 17: 1093–1101.[CrossRef][Medline]
5. Dong J, Calkins H, Solomon SB, Lai S, Dalal D, Lardo AC, Brem E, Preiss A, Berger RD, Halperin H, Dickfeld T. Integrated electroanatomic mapping with three-dimensional computed tomographic images for real-time guided ablations. Circulation. 2006; 113: 186–194.
6. Sra J, Krum D, Hare J, Okerlund D, Thompson H, Vass M, Schweitzer J, Olson E, Foley WD, Akhtar M. Feasibility and validation of registration of three-dimensional left atrial models derived from computed tomography with a noncontact cardiac mapping system. Heart Rhythm. 2005; 2: 55–63.[CrossRef][Medline]
7. Fahmy TS, Mlcochova H, Wazni OM, Patel D, Cihak R, Kanj M, Beheiry S, Burkhardt JD, Dresing T, Hao S, Tchou P, Kautzner J, Schweikert RA, Arruda M, Saliba W, Natale A. Intracardiac echo–guided image integration: optimizing strategies for registration. J Cardiovasc Electrophysiol. 2007; 18: 276–282.[CrossRef][Medline]
8. Zhong H, Lacomis JM, Schwartzman D. On the accuracy of CartoMerge for guiding posterior left atrial ablation in man. Heart Rhythm. 2007; 4: 595–602.[CrossRef][Medline]
9. Bruce GK, Bunch TJ, Milton MA, Sarabanda A, Johnson SB, Packer DL. Discrepancies between catheter tip and tissue temperature in cooled-tip ablation: relevance to guiding left atrial ablation. Circulation. 2005; 112: 954–960.
10. Hocini M, Sanders P, Jais P, Hsu LF, Weerasoriya R, Scavee C, Takahashi Y, Rotter M, Raybaud F, Macle L, Clementy J, Haissaguerre M. Prevalence of pulmonary vein disconnection after anatomical ablation for atrial fibrillation: consequences of wide atrial encircling of the pulmonary veins. Eur Heart J. 2005; 26: 696–704.
11. Packer DL, Stevens CL, Curley MG, Bruce CJ, Miller FA, Khandheria BK, Oh JK, Sinak LJ, Seward JB. Intracardiac phased-array imaging: methods and initial clinical experience with high resolution, under blood visualization: initial experience with intracardiac phased-array ultrasound. J Am Coll Cardiol. 2002; 39: 509–516.
12. Verma A, Marrouche NF, Natale A. Pulmonary vein antrum isolation: intracardiac echocardiography–guided technique. J Cardiovasc Electrophysiol. 2004; 15: 1335–1340.[CrossRef][Medline]
13. Gepstein L, Hayam G, Ben-Haim SA. A novel method for nonfluoroscopic catheter-based electroanatomical mapping of the heart: in vitro and in vivo accuracy results. Circulation. 1997; 95: 1611–1622.
14. Noseworthy PA, Malchano ZJ, Ahmed J, Holmvang G, Ruskin JN, Reddy VY. The impact of respiration on left atrial and pulmonary venous anatomy: implications for image-guided intervention. Heart Rhythm. 2005; 2: 1173–1178.[CrossRef][Medline]
15. Sigfridsson A, Kvitting JP, Knutsson H, Wigstrom L. Five-dimensional MRI incorporating simultaneous resolution of cardiac and respiratory phases for volumetric imaging. J Magn Reson Imaging. 2007; 25: 113–121.[CrossRef][Medline]
16. Hong C, Becker CR, Huber A, Schoepf UJ, Ohnesorge B, Knez A, Brüning R, Reiser MF. ECG-gated reconstructed multi-detector row CT coronary angiography: effect of varying trigger delay on image quality. Radiology. 2001; 220: 712–717.
Related Articles
This article has been cited by other articles:
CLINICAL PERSPECTIVE
During the past 10 years, technology to fully "register" cardiac activation to actual anatomic images has evolved in an effort to facilitate complex anatomy-based ablation. Although these technologies were widely accepted in the electrophysiology field, multiple factors distort actual anatomic information on 3D electroanatomic maps, impedance maps, or merged computed tomographic (CT) image in guiding ablation. To compensate for inaccuracy on these images, this novel 3D ultrasound system was previously developed. This study examined the accuracy and clinical feasibility of guiding complex anatomy-based ablation in animal and human cases. These data demonstrate that nearly real-time acquired 3D ultrasound images can be used to seamlessly generate anatomically accurate chamber geometries. Quantitative information demonstrates minimal error created by the 3D ultrasound registration system, which was more accurate than seen with quantitative merged CT image analysis or standard electroanatomic mapping. This novel system may provide advantages for complex ablation strategies.
Circ Arrhythm Electrophysiol 2008 1: 120-126.
Circ Arrhythm Electrophysiol 2008 1: 103-109.
Circ Arrhythm Electrophysiol 2008 1: 120-126.
![]() |
E. M. Aliot, W. G. Stevenson, J. M. Almendral-Garrote, F. Bogun, C. H. Calkins, E. Delacretaz, P. D. Bella, G. Hindricks, P. Jais, M. E. Josephson, et al. EHRA/HRS Expert Consensus on Catheter Ablation of Ventricular Arrhythmias: Developed in a partnership with the European Heart Rhythm Association (EHRA), a Registered Branch of the European Society of Cardiology (ESC), and the Heart Rhythm Society (HRS); in collaboration with the American College of Cardiology (ACC) and the American Heart Association (AHA) Europace, June 1, 2009; 11(6): 771 - 817. [Full Text] [PDF] |
||||
![]() |
L. F. Tops and J. J. Bax The Year in Imaging Related to Electrophysiology J. Am. Coll. Cardiol. Img., April 1, 2009; 2(4): 498 - 510. [Full Text] [PDF] |
||||
![]() |
Z. M. Hijazi, K. Shivkumar, and D. J. Sahn Intracardiac Echocardiography During Interventional and Electrophysiological Cardiac Catheterization Circulation, February 3, 2009; 119(4): 587 - 596. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Home | Subscriptions | Archives | Feedback | Authors | Help | Circulation Journals Home | AHA Journals Home | Search Copyright © 2008 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |