From the Cardiac Arrhythmia Service (S.M.S., A.d., J.N.R., V.Y.R.) and Gastrointestinal Endoscopy Unit (W.R.B.), Massachusetts General Hospital, Boston, Mass; Cardiac Electrophysiology and Pacing Service, St. Johns Health Center, Santa Monica, Calif (S.K.D.); Southern California Medical Gastroenterology Group, Santa Monica, Calif (R.A.B.).
Correspondence to Andre dAvila, MD, University of Miami Hospital, 1400 NW 12th Ave, Suite 4062, Miami, FL 33136. E-mail adavila{at}med.maimi.edu
Received March 20, 2008; accepted May 30, 2008.
| Abstract |
|---|
|
|
|---|
Methods and Results— Eighty-one consecutive patients who underwent atrial fibrillation ablation followed by esophageal endoscopy were included in this observational study. All patients underwent extraostial electric pulmonary vein isolation by using an electroanatomic mapping system and irrigated RF ablation. All RF applications on the posterior left atrium were limited to 35 W. A commercially available, single-thermocouple esophageal probe was used to monitor LET in a subset of patients (n=67). In these cases, applications were promptly interrupted when LET was
38.5°C; further applications were performed at reduced power to obtain a LET <38.5°C. Esophageal endoscopy was performed 1 to 3 days after the procedure. Ablation-related esophageal ulcerations were identified in 9 of 81 (11%) patients. All patients were asymptomatic. Of these 81 patients, LET monitoring during ablation occurred in 67 (83%) of patients. Esophageal injury was observed more frequently (36% versus 6%, P<0.006) in the group without LET monitoring.
Conclusions— These data suggest that LET monitoring may be associated with a reduction in esophageal injury compared with power limitation alone.
Key Words: catheter ablation fibrillation, atrial complications
| Introduction |
|---|
|
|
|---|
Editorial see p 150
Clinical Perspective see p 162
Left atrial-esophageal (LA-Eso) fistula formation is now a well-recognized complication of percutaneous AF ablation, with an estimated incidence of approximately 0.1%. Although less prevalent than reported in the surgical literature,8 this complication remains devastating, with significant morbidity and mortality.1,4–7 Although the pathophysiology is not fully understood, it is clear that thermal injury to the esophagus during ablation of the LA posterior wall plays a crucial role in triggering the cascade of events that eventually result in the development of LA-Eso fistula.8–10
Currently, the most commonly used clinical strategy to minimize esophageal thermal injury during AF ablation involves limiting the magnitude of power (25 to 35 W), as well as the duration (<30 s), of RF applications placed along the posterior wall of the LA.11–13 A major limitation of this approach is that it fails to account for the variability in the thickness of the posterior LA wall and the presence of periesophageal connective tissue—important determinants of esophageal heating.7,14,15 Thus, empirically limiting the power and duration of RF applications may be insufficient to prevent esophageal thermal injury in all patients. The aim of this study was to determine whether RF power delivery during AF ablation guided by luminal esophageal temperature (LET) monitoring is associated with less frequent esophageal injury compared with a strategy of power limitation alone.
| Methods |
|---|
|
|
|---|
Patient demographics (age, sex, body mass index), disease characteristics (paroxysmal AF, normal heart, ejection fraction, LA size, previous AF ablation), and medication use (current use of antiarrhythmic drugs, aspirin, and proton pump inhibitor or histamine-2 receptor antagonist) were obtained.
Monitoring Esophageal Temperature
In patients who underwent LET monitoring, a 9F single-thermocouple esophageal temperature probe (Vital Temp, Vital Signs Colorado Inc.) was inserted nasally and advanced into the esophagus under fluoroscopic guidance to a position directly posterior to the LA. The position of the tip of the temperature probe was continuously verified by fluoroscopy throughout the procedure and was adjusted in a craniocaudal fashion to the level of the ablation catheter before the application of each ablation lesion to ensure that the thermometer tip was positioned as close as possible to the RF ablation catheter tip (Figure 1). Continuous real-time esophageal temperature monitoring was performed and recorded throughout each case.
|
RF pulses were delivered circumferentially around each vein pair approximately 1 cm from each PV ostium; additional linear lesions were placed, and ablation of fractionated potentials was performed at the operators discretion. RF energy was delivered at the lesions by using a 3.5-mm externally (Thermocool, Biosense Webster) or 4-mm internally (Chili, Boston Scientific) irrigated catheter. The RF generator (Stockert, Biosense Webster) was set to deliver RF energy of up to 35 W and 40°C.
RF applications were terminated when the LET exceeded 38.5°C in patients with LET monitoring. Subsequent lesions in the adjacent areas were ablated with reduced power to avoid further temperature elevation. The maximal LET for each patient was documented with the average power and duration of ablation for the corresponding RF lesion.
Esophageal Evaluation/Follow-Up
All patients underwent a non–symptom-driven EGD between 1 and 3 days after procedure. Ulceration was attributed to AF ablation if it was located on the anterior wall of the midesophagus (approximately 20 to 30 cm from the incisors) and was adjacent to the pulsating heart. These findings were confirmed by the gastroenterologist who performed the EGD (who were blinded to the use of LET monitoring). All patients with ulcerations were treated with high-dose proton-pump inhibitor therapy for at least 1 week, and all underwent repeat endoscopy after 1 week to ensure ulcer healing. Any additional follow-up was scheduled at the discretion of the gastroenterologist.
Typically, the day after the procedure, patients were discharged on anticoagulation with warfarin (target international normalized ratio of 2.0 to 3.0) for at least 3 months after ablation. Preablation antiarrhythmic drugs were generally maintained for 4 weeks after the procedure. Patients were seen in the outpatient clinic for follow-up at 6 weeks and 3 months after procedure. AF recurrences were documented with electrocardiograms, Holter, or transtelephonic monitoring. The presence of esophageal ulceration, postprocedure pericarditis, or pericardial effusion, and AF recurrence at 3 months was ascertained for each patient.
Statistical Analysis
Patients were divided into two groups: those with and without LET monitoring. The primary end point of the study was the presence of esophageal ulceration on EGD. Continuous variables were reported as mean ± SD, and distribution of discrete variables were reported as percentages for each group. Means of continuous variables were compared with the Student t test. Two-by-two contingency tables were created for discrete variables, and frequencies were compared using Fishers exact test when any cell within the table contained a value
5; the
2 test was used in all other cases. The strength and direction of the association between LET and RF power applied were quantified using Pearsons correlation coefficient. Probability values <0.05 were considered significant. All statistical analyses were performed with SAS version 9 (SAS institute, Cary, NC). The authors had full access to the data and take responsibility for its integrity. All authors have read and agree to the manuscript as written.
| Results |
|---|
|
|
|---|
Sixty-seven patients (83%) underwent AF ablation with LET monitoring. Fourteen (17%) did not undergo LET monitoring: five (36%) because of difficulties in placing the LET probe within the esophagus and nine (64%) because of the operators preference. LET monitoring was used with equal frequency throughout the study period and among all operators.
The Table displays the characteristics of the two groups. There were no significant differences between each group, with the exception that individuals without LET monitoring had a lower body mass index, used more antiarrhythmic drugs, and underwent ablation more frequently with general anesthesia.
|
|
|
39°C at least once in more than half of the study cohort (53% [24/45]; maximum recorded LET=41.9°C). There was no clear relationship between the maximal esophageal temperature recorded (Figure 4) or number of elevated LET recording (Figure 5) in the patients with and without esophageal ulceration. A trend toward more esophageal ulceration was noted in patients with maximal LET
39°C (75% versus 54%, P=0.4).
|
|
38.5°C (n=101), 34% occurred at a power of
20 W, including 8% at a power
10 W.
|
| Discussion |
|---|
|
|
|---|
Esophageal Injury Associated With AF Ablation
The mechanism of esophageal injury during RF ablation is not clear but hypothesized to be secondary to thermal injury related to conductive heating.11–13 It has been suggested that the muscular layer of the esophagus absorbs most of the heat and is most susceptible to injury.8,12 Injury to this area may result in reduced esophageal elasticity producing symptoms of dysphagia.4
Aupperle et al12 systematically evaluated esophageal injury in sheep undergoing AF ablation. Twenty-four (62%) sheep demonstrated histological evidence of post-AF ablation esophageal injury, which occurred predominantly in the periesophageal and outer esophageal muscular layers. Only one animal had visible evidence of esophageal injury.
To our knowledge, three human studies have attempted to evaluate the presence of esophageal injury associated with AF ablation. Cummings et al13 performed an EGD on 16 patients who underwent AF ablation with an 8-mm nonirrigated catheter, with power titration guided by microbubbles formation on intracardiac echocardiography. No patient demonstrated esophageal injury on EGD within 24 hours after ablation. Marrouche et al16 examined 28 patients with EGD post-AF ablation (14 underwent AF ablation with an irrigated tip catheter and 14 with an 8-mm nonirrigated catheter). Overall, 13 patients (48%) demonstrated EGD evidence of esophageal injury—5 (18%) developed ulceration and 8 (29%) erythema. Nakagawa et al17 reported EGD findings on 16 patients who underwent AF ablation under general anesthesia with power limitation of 20 to 25 W in areas near the esophagus; nine patients (56%) developed esophageal ulceration. Thus, the rate of AF-related esophageal ulceration during AF ablation is variable.
Use of Esophageal Temperature Monitoring During AF Ablation
Continuous evaluation of the intraluminal esophageal temperature during the ablation procedure may prevent thermal injury. However, this approach has several limitations: (1) accurate intraluminal temperature monitoring depends on optimal contact between the probe and esophageal wall, which is difficult to ensure during the procedure; and (2) intraluminal esophageal temperature underestimates intramural thermal damage to the muscular layers of the esophagus. In addition, the complex anatomy of the esophagus (including the presence of transverse folds) and the malalignment of the LET probe at the region of ablation may result in a slow rise in temperature, with ensuing esophageal injury occurring at otherwise acceptable LETs.
Because of the inherent limitations of LET monitoring, LET was not used to titrate power up in our study. Rather, LET was used simply to ensure that the power settings thought to be safe would not result in undesired high esophageal temperatures. Thus, intraluminal esophageal temperatures were used to interrupt RF applications prematurely if high esophageal temperatures (
38.5°C) were detected at a given power and subsequently titrate power down with additional RF applications. In this regard, this study is important, as it suggests that LET monitoring may confer additional protection from esophageal thermal injury as compared with a strategy of power limitation alone. This is of particular importance because 34 of 101 RF applications with resulting LET
38.5°C occurred despite a reduction in power to
20 W (Figure 6). High LETs were also noted despite power applications of <10 W, suggesting that in certain patients, thermal injury may occur despite significant reductions in RF power. These findings suggest value for LET monitoring during AF ablation.
In patients who underwent LET monitoring, a clear relationship between the maximum LET and number of LET
38.5°C was not observed in those with and without ulceration. Despite this finding, we suspect that the protective benefit of LET monitoring still lies in its ability to detect elevated LET. We hypothesize that empirical ablation of the posterior LA at a constant RF power of 35 W may result in frequent undetected elevations of LET with ensuing esophageal thermal injury, whereas prompt identification of elevated LET during RF ablation in the posterior LA with subsequent down-titration of power to avoid continuous elevated LET may minimize this injury. Although difficult to prove, the presence of longer linear ulcerations in patients without LET monitoring, compared with the discrete ulcerations in patients with LET monitoring, supports this hypothesis.
Advantages and Limitations of LET
Compared with alternative approaches to minimize esophageal injury, LET monitoring is straightforward, inexpensive, well tolerated, and may provide information on the risk of esophageal thermal injury independent of the RF application settings used during ablation.
The optimal LET threshold to avoid injury has not been defined, and must account for limitations in proper LET probe positioning, thermal latency, and the underestimation of mural temperatures with luminal recordings.18 A LET cutoff of 38.5°C was empirically selected in this study based on our prior experience with LET monitoring during AF ablation. Although patients with a LET
39°C had a trend toward ulceration, we are unable to suggest a specific LET cutoff to end ablation lesions, as the maximal LET recorded is frequently higher than the LET at which ablation is terminated. Similar to the conclusions obtained by other investigators,11,13 no relationship between RF power and LET was present.
Theoretically, the use of an LET probe is associated with its own risks, including traumatic insertion (with hematoma formation or perforation), aspiration, and patient intolerance. Although no complications of LET monitoring were noted in this study, probe placement was not possible in five patients. However, with increasing experience, it is possible to properly place the LET probe in most patients.
Titrating power and limiting ablation duration to rises in LET may result in the administration of short, low power RF lesions, which may theoretically reduce the efficacy of AF ablation. This was not supported in this study, as no difference in the rate of AF recurrence was noted at 3 months in both groups. Thus, a strategy of power titration guided by esophageal temperature monitoring does not compromise on efficacy for safety.
Alternative Esophageal Protective Strategies
Alternative strategies described to avoid esophageal injury during AF ablation have included visualization of the esophagus with an enteral feeding tube, barium paste, or electroanatomic mapping,19,20 cooling of the esophagus to counter RF ablation-associated increases in esophageal temperature,21 and using cryoablation rather than RF energy.22 However, the effectiveness of these approaches on minimizing esophageal injury has not been thoroughly studied.
Study Limitations
Our study has several limitations. First, this was an observational, nonrandomized study, and hence, unmeasured confounding factors may be present and not accounted for. Second, the sample size was small and unbalanced between the two groups, thereby reducing the statistical power to ascertain the role of additional clinical and procedural factors (such as the use of general anesthesia) in the development of esophageal injury. However, this is the largest series of its kind assessing the use of LET during AF ablation as well as systematically evaluating for esophageal injury after ablation. Moreover, despite the small sample size, our results are statistically significant. Third, the end point of visible esophageal mucosal injury may result in an underestimation of true incidence of esophageal injury, which may be microscopic12 or limited to the muscular layer.8,12 Estimating the periesophageal and muscular injury to the esophagus can only be adequately performed at autopsy, which is not feasible for a human study of this nature. In addition, we believe that the outcome of interest (ie, visible endoscopic esophageal injury) is clinically relevant. Fourth, the distance between the left atria and the esophagus was not accounted for in this analysis. Although the total tissue thickness between the esophagus and LA is an important factor determining esophageal injury,15 computed tomography analyses have suggested a close relationship of the posterior LA and esophagus in virtually all patients.23 Moreover, adjusting for patients body mass index did not result in any additional protection from esophageal ulceration.
The strengths of this study include its size (which is the largest study of its kind to date), complete follow-up, the simple intervention examined, and the clinically relevant end point reported.
Clinical Implications
Although limiting RF power titration and ablation duration in response to esophageal temperature
38.5°C does not eliminate AF ablation-associated esophageal injury, these data suggest that it may be associated with reduced esophageal injury. As the frequency of LA-eso fistula is approximately 0.1%, it would be extremely difficult to prove that this strategy reduces this complication; however, the reduction of collateral damage to the esophagus remains significant.
| Conclusions |
|---|
|
|
|---|
| Acknowledgments |
|---|
Funding Sources
Dr. S. M. Singh is a recipient of a Detweiler Traveling Fellowship from the Royal College of Physicians and Surgeons of Canada, Ottawa, Ontario, Canada.
Disclosures
None.
| References |
|---|
|
|
|---|
2. Shah D, Dumonceau JM, Burri H, Sunthorn H, Schroft A, Gentil-Baron P, Yokoyama Y, Takahasi A. Acute pyloric spasm and gastric hypomotility: an extracardiac adverse effect of percutaneous radiofrequency ablation for atrial fibrillation. J Am Coll Cardiol. 2005; 46: 327–30.
3. Anfinsen OG, Aass H, Kongsgaard E, Foerster A, Scott H, Amlie JP. Temperature controlled radiofrequency catheter ablation with a 10 mm tip electrode creates larger lesions without charring in the porcine heart. J Interv Card Electrophysiol. 1999; 3: 343–351.[CrossRef][Medline]
4. Scanavacca M, D'Avila A, Parga J, Sosa E. Left atrial-esophageal fistula following radiofrequency catheter ablation of atrial fibrillation. J Cardiovasc Electrophysiol. 2004; 15: 960–962.[Medline]
5. Pappone C, Oral H, Santinelli V, Vicedomini G, Lang CC, Manguso F, Torraca L, Benussi S, Alfieri O, Hong R, Lau W, Hirata K, Shikuma N, Hall B, Morady F. Atrio-esophageal fistula as a complication of percutaneous transcatheter ablation of atrial fibrillation. Circulation. 2004; 109: 2724–2726.
6. Cummings JE, Schwelkert RA, Salida W, Burkhardt D, Kilikaslan F, Saad E, Natale A. Brief communication: atrial-esophageal fistulas after radiofrequency ablation. Ann Intern Med. 2006; 144: 572–574.
7. Ren J, Lin D, Marchlinski F, Callans D, Patel V. Esophageal imaging and strategies for avoiding injury during left atrial ablation for atrial fibrillation. Heart Rhythm. 2006; 3: 1156–1161.[Medline]
8. Doll N, Borger MA, Fabricius A, Stephan S, Gummert J, Mohr FW, Hauss J, Kottkamp H, Hindricks G. Esophageal perforation during left atrial radiofrequency ablation: is the risk too high? J Thorac Cardiovasc Surg. 2003; 125: 836–42.
9. Gillinov AM, Pettersson G, Rice TW. Esophageal injury during radiofrequency ablation for atrial fibrillation. J Thorac Cardiovasc Surg. 2001; 122: 1239–1240.
10. Mohr FW, Fabricius AM, Falk V, Autschbach R, Doll N, Von Oppell U, Diegeler A, Kottkamp H, Hindricks G. Curative treatment of atrial fibrillation with intraoperative radiofrequency ablation: short-term and midterm results. J Thorac Cardiovasc Surg. 2002; 123: 919–927.
11. Redfearn DP, Trim GM, Skanes AC, B, Krahn AD, Yee R, Klein GJ. Esophageal temperature monitoring during radiofrequency ablation of atrial fibrillation. J Cardiovasc Electrophysiol. 2005; 16: 589–593.[CrossRef][Medline]
12. Aupperle H, Doll N, Walther T, Kornherr P, Ullmann C, Schoon H, Mohr FW. Ablation of atrial fibrillation and esophageal injury: effects of energy source and ablation technique. J Thorac Cardiovasc Surg. 2005; 130: 1549–1554.
13. Cummings JE, Schweikert RA, Saliba WI, Burkhardt D, Brachmann J, Gunther J, Schibgilla V, Verma A, Dery M, Drago JL, Kilicaslan F, Natale A. Assessment of temperature, proximity, and course of the esophagus during radiofrequency ablation within the left atrium. Circulation. 2005; 112: 459–464.
14. Sanchez-Quintana D, Cabrera JA, Climent V, Farre J, de Mendonca MC, Ho SY. Anatomic relations between the esophagus and left atrium and relevance for ablation of atrial fibrillation. Circulation. 2005; 112: 1400–1405.
15. Berjano EJ, Hornero F. What affects esophageal injury during radiofrequency ablation of the left atrium? An engineering study based on finite-element analysis. Physiol Meas. 2005; 26: 837–848.[CrossRef][Medline]
16. Marrouche NF, Guenther J, Segerson JM, Daccarett M, Rittger H, Marschang H, Schibgilla V, Schmidt M, Ritscher G, Noelker G, Brachmann J. Randomized comparison between open irrigation technology and intracardiac-echo-guided delivery for pulmonary vein antrum isolation: procedural parameters, outcomes and the effect on esophageal injury. J Cardiovasc Electrophysiol. 2007; 18: 583–588.[CrossRef][Medline]
17. Nakagawa H, Seres KA, Yokoyama K, Collier J, Katari V, Lockwood DJ, Po SS, Slobodova A, Herring L, Olsen J, Lazzara R, Harty RF, Beckman KJ, Jackman WM. High incidence of asymptomatic ulceration after pulmonary vein antrum isolation in patients with atrial fibrillation [Abstract]. Heart Rhythm. 2007; 4 (suppl): S61.[CrossRef][Medline]
18. Perzanowski C, Teplitsky L, Hrantizky PM, Bahnson TD. Real-time monitoring of luminal esophageal temperature during left atrial radiofrequency catheter ablation for atrial fibrillation: observations about heating during ablation at the pulmonary vein ostia and left atrium. J Cardiovasc Electrophysiol. 2006; 17: 166–170.[CrossRef][Medline]
19. Hall B, Shah A, Huang D, Rosero S, Daubert J. Visualization of the esophagus during catheter ablation of atrial fibrillation. J Interv Card Electophysiol. 2005; 13: 135–137.[CrossRef]
20. Pollack SJ, Monir G, Chernoby MS, Ellenberger CD. Novel imaging techniques of the esophagus enhancing safety of left atrial ablation. J Cardiovasc Electrophysiol. 2005; 16: 244–248.[CrossRef][Medline]
21. Tsuchiya T, Ashikaga K, Nakagawa S, Hayashida K, Kugimiya H. Atrial fibrillation ablation with esophageal cooling with a cooled-water irrigated intraesophageal balloon. J Cardiovasc Electrophysiol. 2007; 18: 145–150.[CrossRef][Medline]
22. Ripley KL, Gage AA, Olsen DB, Van Vleet JF, Lau CP, Tse HF. Time course of esophageal lesions after catheter ablation with cryothermal and radiofrequency ablation: implications for atrio-esophageal fistula formation after catheter ablation for atrial fibrillation. J Cardiovasc Electrophysiol. 2007; 18: 642–646.[CrossRef][Medline]
23. Cury RC, Abbara S, Schmidt S, Malchano ZJ, Neuzil P, Weichet J, Ferenick M, Hoffman U, Ruskin JN, Reddy VY. Relationship of the esophagus and aorta to the left atrium and pulmonary veins: implications for catheter ablation of atrial fibrillation. Heart Rhythm. 2005; 2: 1317–1223.[CrossRef][Medline]
| Footnotes |
|---|
Related Articles
Circ Arrhythm Electrophysiol 2008 1: 150-152.
Circ Arrhythm Electrophysiol 2008 1: 162-168.
This article has been cited by other articles:
![]() |
T. J. Badger, Y. A. Adjei-Poku, N. S. Burgon, S. Kalvaitis, A. Shaaban, D. N. Sommers, J. J.E. Blauer, E. N. Fish;, N. Akoum, T. S. Haslem, et al. Initial Experience of Assessing Esophageal Tissue Injury and Recovery Using Delayed-Enhancement MRI After Atrial Fibrillation Ablation Circ Arrhythm Electrophysiol, December 1, 2009; 2(6): 620 - 625. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Sohara, H. Takeda, H. Ueno, T. Oda, and S. Satake Feasibility of the Radiofrequency Hot Balloon Catheter for Isolation of the Posterior Left Atrium and Pulmonary Veins for the Treatment of Atrial Fibrillation Circ Arrhythm Electrophysiol, June 1, 2009; 2(3): 225 - 232. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Di Biase, L. C. Saenz, D. J. Burkhardt, M. Vacca, C. S. Elayi, C. D. Barrett, R. Horton, R. Bai, A. Siu, T. S. Fahmy, et al. Esophageal Capsule Endoscopy After Radiofrequency Catheter Ablation for Atrial Fibrillation: Documented Higher Risk of Luminal Esophageal Damage With General Anesthesia as Compared With Conscious Sedation Circ Arrhythm Electrophysiol, April 1, 2009; 2(2): 108 - 112. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Nakagawa, K. A. Seres, and W. M. Jackman Limitations of Esophageal Temperature-Monitoring to Prevent Esophageal Injury During Atrial Fibrillation Ablation Circ Arrhythm Electrophysiol, August 1, 2008; 1(3): 150 - 152. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Home | Subscriptions | Archives | Feedback | Authors | Help | Circulation Journals Home | AHA Journals Home | Search Copyright © 2008 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |