Original Articles |
From the Canadian Adult Congenital Heart Network, Montreal, Quebec, Canada (P.K., L.H., A.B., L.A.M., P.C., E.G., A.D.); Leeds General Infirmary, Leeds, United Kingdom (S.V., P.C.); and Childrens Hospital, Boston, Mass (P.K., M.J.L., S.M.F., F.C.).
Correspondence to Dr Paul Khairy, Adult Congenital Heart Center, Montreal Heart Institute, 5000 Bélanger Street, Montreal, Quebec, Canada H1T 1C8. E-mail paul.khairy{at}umontreal.ca
Received February 26, 2008; accepted July 23, 2008.
| Abstract |
|---|
|
|
|---|
Methods and Results— We conducted a multicenter cohort study in patients with transposition of the great arteries to determine actuarial rates of implantable cardioverter defibrillator shocks, identify risk factors, assess underlying arrhythmias, and characterize complications. Overall, 37 patients (age, 28.0±7.6 years; 89.2% male) were enrolled from 7 sites. Implantable cardioverter defibrillators were implanted for primary prevention in 23 (62.1%) patients and secondary prevention in 14 patients (37.8%). Annual rates of appropriate shocks were 0.5% and 6.0% in primary and secondary prevention, respectively (P=0.0366). Independent predictors were a secondary prevention indication (hazard ratio, 18.0; P=0.0341) and lack of β-blockers (hazard ratio, 16.7; P=0.0301). In patients with appropriate shocks, intracardiac electrograms documented supraventricular tachycardia preceding or coexisting with ventricular tachycardia in 50%. No patient with inducible ventricular tachycardia received an appropriate shock in comparison with 37.5% of noninducible patients (P=0.0429). Inappropriate shocks occurred in 6.6% per year, more so in patients of lesser weight (hazard ratio, 0.91 per kg; P=0.0168). Additionally, 14 patients (37.8%) experienced complications: 5 (13.5%) acute, 1 (2.7%) late generator related, and 12 (32.4%) late lead related.
Conclusion— In patients with transposition of the great arteries, high rates of appropriate shocks are noted in secondary but not primary prevention. Supraventricular arrhythmias may be implicated in the etiology of ventricular tachyarrhythmias; β-blockers seem protective, and inducible ventricular tachycardia does not seem to predict future events. Inappropriate shocks and late lead-related complications are common.
Key Words: death, sudden heart arrest defibrillation heart defects, congenital tachyarrhythmias transposition of great vessels
| Introduction |
|---|
|
|
|---|
Clinical Perspective see p 250
Experience with implantable cardioverter defibrillators (ICDs) in patients with D-TGA is limited, whether prompted by a sustained ventricular tachyarrhythmia or resuscitated cardiac arrest (secondary prevention indication), or a clinical profile deemed high risk in the absence of a near fatal event (primary prevention indication). We conducted a multicenter retrospective cohort study with the objectives of determining actuarial rates of appropriate and inappropriate ICD shocks in patients with D-TGA, identifying associated risk factors, and characterizing complications. It was also our hope that analysis of ICD electrogram recordings would provide mechanistic insights underlying sudden deaths.
| Methods |
|---|
|
|
|---|
Baseline Characteristics
A similar protocol was previously described in patients with tetralogy of Fallot.9 Data collection was conducted in accordance with individual hospital institutional review board policies. Details regarding demographic variables, surgical history including type of intra-atrial baffle repair (ie, Mustard or Senning), associated anomalies, interventions before or concomitant with surgical repair, electrocardiography, chest radiography, 24-hour Holter monitoring, 2D and M-mode Doppler echocardiography, cardiac magnetic resonance (CMR) imaging, cardiac catheterization, and electrophysiology studies were collected. The most recent data preceding ICD implantation were requested, with a maximum acceptable time interval of 2 years.
Electrocardiographic data included heart rate, presence of an underlying ventricular-paced rhythm, longest dominant QRS duration, and QT intervals. Cardiothoracic ratios were derived from posterioanterior chest radiographs. The number of premature ventricular complexes in 24 hours and presence of nonsustained ventricular tachycardia (
3 beats, <30 seconds) were captured from Holter monitors. Echocardiographic and CMR data included biventricular size and function, degree of valvar regurgitation, and estimates of systolic pulmonary arterial pressure. When both echocardiographic and CMR studies were performed, CMR data were retained. Data extracted from programmed ventricular stimulation studies included inducibility of sustained monomorphic or polymorphic ventricular tachycardia.
ICD Implantation
Indications prompting ICD implantation were recorded and included presyncope/dizziness, syncope, palpitations, clinical nonsustained ventricular tachycardia, QRS duration
180 ms, severe systemic ventricular systolic dysfunction, inducible ventricular tachycardia, clinical sustained ventricular tachycardia, ventricular fibrillation/resuscitated cardiac arrest, and other. Patients were stratified according to whether the ICD was indicated for primary or secondary prevention. "Secondary prevention" was defined by clinical sustained ventricular tachycardia, ventricular fibrillation, or resuscitated cardiac arrest. As per convention, barring such events, the ICD was considered indicated for "primary prevention."
Procedural characteristics were logged and medical therapy at the time of discharge was noted, with particular attention to β-blockers, amiodarone, sotalol, dofetilide, and class IA or IC antiarrhythmic agents.
ICD Shocks
The main outcome consisted of appropriate ICD shocks. The ventricular tachycardia cycle length, programmed defibrillator zone (ie, slow ventricular tachycardia, fast ventricular tachycardia, or ventricular fibrillation), and success or failure of therapy was noted. Data regarding inappropriate ICD shocks were likewise collected. All ICD events (ie, antitachycardia pacing or shock) and tracings were requested, up to a maximum of 5 per "appropriate" and "inappropriate" category for each patient.
A blinded adjudicating committee reviewed and classified all ICD events, with each tracing analyzed by
2 independent electrophysiologists. Appropriate therapy was subclassified as monomorphic ventricular tachycardia (ie, electrogram with a uniform and constant morphology), polymorphic ventricular tachycardia (ie, electrogram with relatively constant amplitude but displaying a shift in morphology or axis), or ventricular fibrillation (ie, constant shift in axis and morphology of the electrogram accompanied by marked and variable changes in amplitude). Inappropriate ICD shocks were further categorized according to type of most likely underlying rhythm (ie, noise interference or oversensing, sinus tachycardia, atrial flutter, atrial fibrillation, and other form of supraventricular tachycardia).
ICD Complications
ICD complications were considered "periprocedural" if they occurred within 30 days of implantation and "late" thereafter. Late complications were subdivided into lead and generator-related events. Lead-related complications included lead dislodgement, lead failure, endocarditis, and undersensing or oversensing. Pain, erosion, pocket infection, migration, and device malfunction were considered generator-related complications.
Statistical Analysis
Time zero was defined as time of ICD implantation. Patient-years were accrued from time of entry until occurrence of an ICD shock or the study termination date. Censoring occurred in the event of cardiac transplantation, loss to follow-up, or death from other causes not involving ICD therapy. Continuous variables are summarized by mean±SD or median and interquartile range (25th, 75th percentile), depending on normality of distribution. Categorical variables are represented by frequencies and percentages. Baseline comparisons between patients with ICDs for primary versus secondary prevention were performed by Mann-Whitney rank sum, student t, or
2 tests where appropriate. Freedom from appropriate and inappropriate ICD shocks and overall survival was plotted using the Kaplan-Meier method, with comparisons by log-rank statistics.
To assess predictors of ICD shocks, univariate and stepwise multivariate Cox proportional hazard models were used after verifying proportional-hazards assumptions. Variables with probability values <0.1 in univariate analyses were considered in multivariate models. Two-tailed probability values <0.05 were considered statistically significant. Analyses were performed with SAS version 9.1 (SAS Institute, Cary, NC).
The authors had full access to and take full responsibility for the integrity of the data. All authors have read and agreed to the manuscript as written.
| Results |
|---|
|
|
|---|
|
|
35% in 8 patients (34.8%), QRS duration
180 ms in 7 patients (30.4%), and inducible sustained ventricular tachycardia in 7 patients (30.4%). Indications for secondary prevention were ventricular fibrillation and/or resuscitated cardiac arrest in 10 patients (71.4%) and clinical sustained ventricular tachycardia without cardiac arrest in 4 patients (28.6%).
Appropriate Antitachycardia Pacing or Shocks
Overall, 139 ICD discharges occurred in 12 patients over a median follow-up of 3.6 (1.5, 5.5) years: 3.5 (1.4, 5.3) years and 3.9 (2.3, 6.6) years in patients with primary and secondary prevention indications, respectively. Of all first ICD discharges, 33.3% were appropriate. Five patients (13.5%) received a total of 47 appropriate shocks (1 in primary prevention; 46 in secondary prevention) clustered as follows: 1, 2, 4, 5, and 35 shocks. The median detected ventricular tachyarrhythmia rate at time of first appropriate shock was 222 (165, 261) bpm. Monomorphic ventricular tachycardia constituted 48.9% of appropriately treated arrhythmias, polymorphic ventricular tachycardia constituted 34.0%, and ventricular fibrillation constituted 17.0%. First ICD discharges consisted of antitachycardia pacing alone in 1, antitachycardia pacing followed by a shock in 1, and shocks alone in 3 patients. The patient with initial antitachycardia pacing alone later received an appropriate ICD shock.
Appropriate ICD Shocks
High rates of appropriate ICD shocks were noted for secondary (28.6%) but not primary (4.3%) prevention indications, with annualized actuarial event rates of 6.0% and 0.5%, respectively (P=0.0366). The only primary prevention ICD patient to receive an appropriate shock was a 33-year-old man with a Mustard baffle, severe systemic right ventricular dysfunction (ejection fraction 12%), moderate tricuspid regurgitation, and nonsustained ventricular tachycardia. Freedom from appropriate ICD shocks in primary and secondary prevention is plotted in Figure 2A. All first appropriate ICD shocks were successful. Univariate and multivariate predictors of appropriate ICD shocks are listed in Table 2. Figure 3 depicts freedom from appropriate ICD shocks according to whether or not patients received β-blockers. Class III antiarrhythmic agents did not significantly modulate risk of appropriate shocks.
|
|
|
Inducible Ventricular Tachycardia and Appropriate ICD Shocks
Of 17 patients with electrophysiological studies, 10 (58.8%) had primary and 7 (41.2%) had secondary prevention indications. Overall, 9 patients (52.9%; 6 primary prevention) had inducible sustained ventricular tachycardia, subclassified as monomorphic in 3 patients (2 primary prevention) and polymorphic in 6 patients (4 primary prevention). Appropriate ICD shocks were received by 0 of 9 patients with inducible sustained ventricular tachycardia and 3 of 8 (ie, 0 of 4 and 3 of 4 with primary and secondary prevention indications, respectively) noninducible patients (37.5%; P=0.0429). Two of the 3 noninducible patients with appropriate ICD shocks received them for monomorphic ventricular tachycardia, whereas the remaining patient had recurrent polymorphic ventricular tachycardia.
Inappropriate ICD Shocks
Overall, 9 patients (24.3%) received 92 inappropriate ICD shocks, 67 of which were in 5 patients (21.7%) with primary prevention and 25 of which were in 4 patients (28.6%) with secondary prevention indications. Underlying causes were noted for 78 inappropriate shocks: lead fracture, failure, or oversensing in 48 patients (61.5%), supraventricular tachycardia in 21 patients (26.9%), and sinus tachycardia in 9 patients (11.5%). The 21 inappropriate shocks for supraventricular arrhythmias occurred in 7 patients and were all designated "atrial flutter." The median detected ventricular tachycardia rate was 188 (165, 200) bpm, significantly slower than appropriately shocked arrhythmias (P=0.0078).
Freedom from inappropriate ICD shocks at 1, 2, and 5 years was 83.2%, 75.6%, and 71.1%, respectively, corresponding to an average actuarial rate of 6.6% per year. No difference was observed between primary and secondary prevention groups (P=0.9765). In univariate analyses, risk factors for inappropriate shocks were female sex (hazard ratio, 4.9; 95% CI, 1.2 to 19.6; P=0.0261) and lighter weight (hazard ratio, 0.91 per kg; 95% CI, 0.85 to 0.98; P=0.0106). In multivariate analyses, the latter was the only independent predictor of inappropriate ICD shocks (hazard ratio, 0.91 per kg; 95% CI, 0.85 to 0.98; P=0.0168).
Complications
Complications other than inappropriate shocks occurred in 14 patients (37.8%) overall: acute in 5 (13.5%), late generator related in 1 (2.7%), and late lead related in 12 (32.4%). Table 3 summarizes complications according to subtype; the categories are not mutually exclusive. Three patients had both acute and late complications.
|
The only episode of sudden death occurred in a 22-year-old man with a Senning baffle, an ICD for secondary prevention, and a prior appropriate and effective shock. His fatal event was triggered by polymorphic ventricular tachycardia that prompted 4 appropriate ICD shocks. Although the final shock successfully interrupted ventricular fibrillation, electromechanical dissociation ensued with an agonal escape rhythm.
| Discussion |
|---|
|
|
|---|
Little is known about ICDs in this patient population, with the literature largely limited to case reports17 and 2 small case series.18,19 One series of 5 patients reported no appropriate shocks at a median of 20 months.18 The second series included 7 patients with D-TGA as part of a larger cohort of adults with congenital heart disease.19 Subgroup analyses specific to D-TGA were not reported. To our knowledge, our multicenter study is the first to explicitly address the role of ICDs in high-risk patients with D-TGA. Importantly, a high rate of appropriate shocks was observed in patients with secondary prevention indications. Ventricular tachyarrhythmias were reliably sensed and effectively interrupted. In contrast, the rate of appropriate ICD shocks in patients with primary prevention indications was exceedingly low, highlighting our limitations in reliably identifying appropriate high-risk candidates. Consequently, prophylactic ICDs in D-TGA for indications such as systemic right ventricular dysfunction, nonsustained ventricular tachycardia, syncope, and inducible ventricular tachycardia remain unsubstantiated.
Little is known about the pathophysiology of sudden death in patients with D-TGA and intra-atrial baffles. To this regard, our study design provided a unique opportunity to clarify mechanistic triggers. Intracardiac ICD tracings were requested on all patients with shocks and subject to systematic blinded adjudication. Interestingly, supraventricular tachyarrhythmias were often found to precede or coexist with ventricular tachycardia. It can be conjectured, therefore, that supraventricular tachyarrhythmias may be an important contributor to sudden death in this patient population and that benefit may be derived from the suppression or elimination of these substrates. This notion is consistent with reported risk factors for sudden death. In a case-control study, Kammeraad et al8 found that arrhythmia symptoms and documented atrial fibrillation or flutter were most predictive of sudden death. Several cohort studies likewise reported an increased incidence of sudden death in patients with supraventricular arrhythmias, although it remained unclear whether these were causative or surrogate markers for structural or hemodynamic impairment.2,20–22
Factors that may enhance susceptibility to degeneration of supraventricular arrhythmias into fatal events include impaired atrioventricular transport with failure to augment right ventricular filling rates during tachycardia,23 atrioventricular nodal conduction capable of rapid ventricular responses,24 systemic right ventricular dysfunction,25 and subendocarial ischemia resulting from a right coronary circulation irrigating a systemic ventricle.26 In addition to potentially suppressing primary ventricular arrhythmias, β-blockers may have salutary effects on a combination of these factors, which may explain, in part, our finding that β-blockers afford protection against appropriate shocks. β-blockers may suppress supraventricular arrhythmias, increase diastolic filling time, slow ventricular responses by increasing the atrioventricular nodal effective refractory period, and reduce myocardial ischemia. Unlike the lack of benefit with angiotensin receptor blockade,27 preliminary data suggests that carvedilol may be associated with positive right ventricular remodeling in patients with systemic right ventricles.28
Despite a predominance of supraventricular arrhythmias, we also observed primary ventricular tachyarrhythmias. Importantly, these occurred exclusively in patients with secondary prevention indications. In a manner similar to dilated29 and hypertrophic30 cardiomyopathy, programmed ventricular stimulation was of limited prognostic value. Fewer appropriate shocks occurred in inducible patients. Although this likely reflects a chance finding, it is improbable that a larger sample size would detect a statistically significant association in the opposite direction. This is unlike tetralogy of Fallot where inducible sustained ventricular tachycardia is of value in predicting clinical ventricular tachyarrhythmia, sudden death, and appropriate shocks.9,31
As for complications, in patients with congenital heart disease, high rates of inappropriate ICD shocks are consistently reported. This likely reflects the relatively young age of the study population, propensity for coexisting supraventricular tachyarrhythmias, and higher susceptibility for lead failure.32 Indeed, nearly one third of our patients experienced late lead-related complications. Prior studies have likewise noted such high rates of complications in patients with congenital heart disease with pacemakers and ICDs.9,19,32–36 Although we found lighter weight to independently predict inappropriate shocks, younger age was previously identified as a risk factor for lead fracture.32
Although some overlap was noted, supraventricular arrhythmias that provoked inappropriate shocks were generally slower than ventricular tachyarrhythmias. Post hoc analysis suggests that inappropriate shocks could have largely been prevented by programming the ventricular tachycardia detection rate >200 bpm. The desirability of such an approach remains speculative, as it cannot be conclusively determined whether inappropriate shocks for supraventricular tachyarrhythmias may have circumvented otherwise fatal events (ie, appropriate "inappropriate" shocks).
Limitations
Our study was retrospective, detection rates for ICD shocks were not uniformly programmed, and medical decisions were not randomly allocated. Primary prevention indications were recorded and assessed, but standardized selection criteria were not applied. Despite being the largest study of its kind, the ability to identify relevant predictors is limited by the overall low event rate. As such, a conservative probability value cutoff (<0.1) was selected to limit potential variables considered in multivariate models. Potential for human error remains despite efforts to minimize misclassification of appropriate versus inappropriate ICD shocks through systematic appraisal by a blinded expert adjudicating committee. In 28% of cases, appropriate ICD shocks were confirmed but provided data were insufficient to comment on preceding supraventricular tachyarrhythmias. Importantly, ICD recipients were not compared to patients with similarly high-risk features but without defibrillators, precluding quantification of survival benefits.
| Conclusion |
|---|
|
|
|---|
| Acknowledgments |
|---|
Sources of Funding
This study was funded by the Alliance for Congenital Heart Quebec Interinstitutional Research, Fonds de Recherce en Santé du Québec, and the Dunlevie Foundation. Dr Khairy is supported by a Canada Research Chair in Electrophysiology and Adult Congenital Heart Disease.
Disclosures
None.
| References |
|---|
|
|
|---|
2. Janousek J, Paul T, Luhmer I, Wilken M, Hruda J, Kallfelz HC. Atrial baffle procedures for complete transposition of the great arteries: natural course of sinus node dysfunction and risk factors for dysrhythmias and sudden death. Z Kardiol. 1994; 83: 933–938.[Medline]
3. Gatzoulis MA, Walters J, McLaughlin PR, Merchant N, Webb GD, Liu P. Late arrhythmia in adults with the mustard procedure for transposition of great arteries: a surrogate marker for right ventricular dysfunction? Heart. 2000; 84: 409–415.
4. Gelatt M, Hamilton RM, McCrindle BW, Connelly M, Davis A, Harris L, Gow RM, Williams WG, Trusler GA, Freedom RM. Arrhythmia and mortality after the Mustard procedure: a 30-year single-center experience. J Am Coll Cardiol. 1997; 29: 194–201.[Abstract]
5. Stark J, Tynan MJ, Ashcraft KW, Aberdeen E, Waterston DJ. Obstruction of pulmonary veins and superior vena cava after the Mustard operation for transposition of the great arteries. Circulation. 1972; 45: I116–I120.[Medline]
6. Benatar A, Tanke R, Roef M, Meyboom EJ, Van de Wal HJ. Mid-term results of the modified Senning operation for cavopulmonary connection with autologous tissue. Eur J Cardiothorac Surg. 1995; 9: 320–324.[Abstract]
7. Kirjavainen M, Happonen JM, Louhimo I. Late results of Senning operation. J Thorac Cardiovasc Surg. 1999; 117: 488–495.
8. Kammeraad JA, van Deurzen CH, Sreeram N, Bink-Boelkens MT, Ottenkamp J, Helbing WA, Lam J, Sobotka-Plojhar MA, Daniels O, Balaji S. Predictors of sudden cardiac death after Mustard or Senning repair for transposition of the great arteries. J Am Coll Cardiol. 2004; 44: 1095–1102.
9. Khairy P, Harris L, Landzberg MJ, Viswanathan S, Barlow A, Gatzoulis MA, Fernandes SM, Beauchesne L, Therrien J, Chetaille P, Gordon E, Vonder Muhll I, Cecchin F. Implantable cardioverter-defibrillators in tetralogy of Fallot. Circulation. 2008; 117: 363–370.
10. Brickner ME, Hillis LD, Lange RA. Congenital heart disease in adults. Second of two parts. N Engl J Med. 2000; 342: 334–342.
11. Senning A. Surgical correction of transposition of the great vessels. Surgery. 1959; 45: 966–980.[Medline]
12. Mustard WT. Successful two stage correction of transposition of the great vessels. Surgery. 1964; 55: 469–472.[Medline]
13. Khairy P, Landzberg MJ, Lambert J, ODonnell CP. Long-term outcomes after the atrial switch for surgical correction of transposition: a meta-analysis comparing the Mustard and Senning procedures. Cardiol Young. 2004; 14: 284–292.[CrossRef][Medline]
14. Khairy P, Dore A, Talajic M, Dubuc M, Poirier N, Roy D, Mercier LA. Arrhythmias in adult congenital heart disease. Expert Rev Cardiovasc Ther. 2006; 4: 83–95.[CrossRef][Medline]
15. Gatzoulis MA, Balaji S, Webber SA, Siu SC, Hokanson JS, Poile C, Rosenthal M, Nakazawa M, Moller JH, Gillette PC, Webb GD, Redington AN. Risk factors for arrhythmia and sudden cardiac death late after repair of tetralogy of Fallot: a multicentre study. Lancet. 2000; 356: 975–981.[CrossRef][Medline]
16. Gewillig M, Cullen S, Mertens B, Lesaffre E, Deanfield J. Risk factors for arrhythmia and death after Mustard operation for simple transposition of the great arteries. Circulation. 1991; 84: III187–III192.[Medline]
17. Lopez JA, Lufschanowski R. Use of a transvenous dual-chamber ICD after a mustard operation for d-transposition of the great vessels. Tex Heart Inst J. 2007; 34: 218–221.[Medline]
18. Michael KA, Veldtman GR, Paisey JR, Yue AM, Robinson S, Allen S, Sunni NS, Kiesewetter C, Salmon T, Roberts PR, Morgan JM. Cardiac defibrillation therapy for at risk patients with systemic right ventricular dysfunction secondary to atrial redirection surgery for dextro-transposition of the great arteries. Europace. 2007; 9: 281–284.
19. Yap SC, Roos-Hesselink JW, Hoendermis ES, Budts W, Vliegen HW, Mulder BJ, van Dijk AP, Schalij MJ, Drenthen W. Outcome of implantable cardioverter defibrillators in adults with congenital heart disease: a multi-centre study. Eur Heart J. 2007; 28: 1854–1861.
20. Flinn CJ, Wolff GS, Dick M, Campbell RM, Borkat G, Casta A, Hordof A, Hougen TJ, Kavey RE, Kugler J. Cardiac rhythm after the Mustard operation for complete transposition of the great arteries. N Engl J Med. 1984; 310: 1635–1638.[Abstract]
21. Sarkar D, Bull C, Yates R, Wright D, Cullen S, Gewillig M, Clayton R, Tunstill A, Deanfield J. Comparison of long-term outcomes of atrial repair of simple transposition with implications for a late arterial switch strategy. Circulation. 1999; 100: II176–II181.[Medline]
22. Woo A, Monakier D, Harris L, Hill A, Shah P, Wigle ED, Rakowski H, Rozenblyum E, Cameron DA. Determinants of implantable defibrillator discharges in high-risk patients with hypertrophic cardiomyopathy. Heart. 2007; 93: 1044–1045.
23. Derrick GP, Narang I, White PA, Kelleher A, Bush A, Penny DJ, Redington AN. Failure of stroke volume augmentation during exercise and dobutamine stress is unrelated to load-independent indexes of right ventricular performance after the Mustard operation. Circulation. 2000; 102: III154–III159.[Medline]
24. Haemmerli M, Bolens M, Friedli B. Electrophysiological studies after the Mustard and Senning operations for complete transposition. Do they have prognostic value? Int J Cardiol. 1990; 27: 167–173.[CrossRef][Medline]
25. Pettersen E, Helle-Valle T, Edvardsen T, Lindberg H, Smith HJ, Smevik B, Smiseth OA, Andersen K. Contraction pattern of the systemic right ventricle shift from longitudinal to circumferential shortening and absent global ventricular torsion. J Am Coll Cardiol. 2007; 49: 2450–2456.
26. Millane T, Bernard EJ, Jaeggi E, Howman-Giles RB, Uren RF, Cartmill TB, Hawker RE, Celermajer DS. Role of ischemia and infarction in late right ventricular dysfunction after atrial repair of transposition of the great arteries. J Am Coll Cardiol. 2000; 35: 1661–1668.
27. Dore A, Houde C, Chan KL, Ducharme A, Khairy P, Juneau M, Marcotte F, Mercier LA. Angiotensin receptor blockade and exercise capacity in adults with systemic right ventricles: a multicenter, randomized, placebo-controlled clinical trial. Circulation. 2005; 112: 2411–2416.
28. Giardini A, Lovato L, Donti A, Formigari R, Gargiulo G, Picchio FM, Fattori R. A pilot study on the effects of carvedilol on right ventricular remodelling and exercise tolerance in patients with systemic right ventricle. Int J Cardiol. 2006.
29. Grimm W, Hoffmann J, Menz V, Luck K, Maisch B. Programmed ventricular stimulation for arrhythmia risk prediction in patients with idiopathic dilated cardiomyopathy and nonsustained ventricular tachycardia. J Am Coll Cardiol. 1998; 32: 739–745.
30. Behr ER, Elliott P, McKenna WJ. Role of invasive EP testing in the evaluation and management of hypertrophic cardiomyopathy. Card Electrophysiol Rev. 2002; 6: 482–486.[CrossRef][Medline]
31. Khairy P, Landzberg MJ, Gatzoulis MA, Lucron H, Lambert J, Marcon F, Alexander ME, Walsh EP. Value of programmed ventricular stimulation after tetralogy of fallot repair: a multicenter study. Circulation. 2004; 109: 1994–2000.
32. Fortescue EB, Berul CI, Cecchin F, Walsh EP, Triedman JK, Alexander ME. Patient, procedural, and hardware factors associated with pacemaker lead failures in pediatrics and congenital heart disease. Heart Rhythm. 2004; 1: 150–159.[CrossRef][Medline]
33. Dore A, Santagata P, Dubuc M, Mercier LA. Implantable cardioverter defibrillators in adults with congenital heart disease: a single center experience. Pacing Clin Electrophysiol. 2004; 27: 47–51.[CrossRef][Medline]
34. Alexander ME, Cecchin F, Walsh EP, Triedman JK, Bevilacqua LM, Berul CI. Implications of implantable cardioverter defibrillator therapy in congenital heart disease and pediatrics. J Cardiovasc Electrophysiol. 2004; 15: 72–76.[Medline]
35. Khairy P, Roux JF, Dubuc M, Thibault B, Guerra PG, Macle L, Mercier LA, Dore A, Roy D, Talajic M, Page P. Laser lead extraction in adult congenital heart disease. J Cardiovasc Electrophysiol. 2007; 18: 507–511.[CrossRef][Medline]
36. Walker F, Siu SC, Woods S, Cameron DA, Webb GD, Harris L. Long-term outcomes of cardiac pacing in adults with congenital heart disease. J Am Coll Cardiol. 2004; 43: 1894–1901.
Related Article
CLINICAL PERSPECTIVE
Transposition of the great arteries with Mustard or Senning baffles is among the congenital heart defects at highest risk for sudden cardiac death. We conducted a multicenter cohort study to determine rates of appropriate and inappropriate ICD shocks, identify risk factors, characterize complications, and elucidate potential mechanistic triggers for sudden deaths. Overall, 37 patients (age, 28.0±7.6 years) were enrolled from 7 sites with ICDs for primary prevention in 23 and secondary prevention in 14. A high rate of appropriate shocks was noted in secondary (6.0% per year) but not primary (0.5% per year) prevention. Independent risk factors were limited to a secondary prevention indication (hazard ratio, 18.0) and lack of β-blockers (hazard ratio, 16.7). Inducible ventricular tachycardia did not seem to predict appropriate shocks. In 50% of cases, intracardiac electrogram tracings recorded supraventricular tachycardia preceding or coexisting with ventricular tachycardia. Inappropriate shocks occurred in 6.6% per year, and 37.8% of patients experienced complications, predominantly late lead related. Although these results generally support the use of ICDs for secondary prevention, they likewise highlight our limitations in reliably identifying appropriate candidates for prophylactic ICDs. Indications such as systemic right ventricular dysfunction, nonsustained ventricular tachycardia, syncope, and inducible ventricular tachycardia remain unsubstantiated. Importantly, it is unknown whether inappropriate shocks for supraventricular tachyarrhythmias may have prevented fatal events (ie, appropriate "inappropriate" shocks). As supraventricular tachyarrhythmias seem to be important contributors to sudden death in this patient population, it may be hypothesized that benefit may be derived from suppressing or eliminating these substrates.
Circ Arrhythm Electrophysiol 2008 1: 250-257.
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|